DT 200 Users Manual version 1.E Set Operation Parameters
25
Parameter 25 Serial Com. Port 2 Mode (4 [8 data bits, no parity])
Enter the mode number from the table below to select the mode parameter in the mode
description column.
Mode Number
Mode Description
1 7 data bits, no parity
2 7 data bits, even parity
3 7 data bits, odd parity
4 8 data bits, no parity
5 8 data bits, even parity
6 8 data bits, odd parity
Parameter 26 Serial Com. Port 2 Baud Rate (1 [9600 baud])
Enter a baud number from the table below to select a baud rate from the description
column. (*HS baud rates for the 100 samples per sec version only).
Baud Number Baud Description
1 9600 Baud
2 4800 Baud
3 2400 Baud
4 1200 Baud
5 600 Baud
6 300 Baud (* HS 19.2K)
7 150 Baud (*HS 38.4K)
Parameter 27 TX2 Control / Station ID (255 [single station])
Serial Port 2 can be used for continuous transmission of scale data or as a network buss
or for single station remote control.
•
Enter a 0 to enable continuous transmission of TX 2 data. The data sent is configured
in parameter 28 (TX2 Format).
•
Enter a station ID (1 to 254) for multi-drop network configuration
•
Enter 255 for single station remote control
Parameter 28 TX2 Format (2 Condec format)
Serial Port 2 can be used for continuous transmission of scale data. The TX2 Format
parameter determines what is sent. There are 5 selections:
1 - AND: <status>, <mode>, <polarity><weight><units><terminate>
Status: OL (overload), ST (stable), US (unstable, in motion)
Mode: GS (gross mode), NT (net mode)
Polarity: + (positive weight), - (negative weight)
Weight: 7 characters data, NOT zero blanked, with decimal point
Units: lb, kg, t (ton), tn (tonne), g<sp> (gram), oz
Terminate: lf, cr (line feed, carriage return)