
DT220 Users Manual version 1.a Configuration Parameters
32
System initialize, save Cal.
Use this function to reset DT220 parameters to factory default values. Calibration factors
are NOT changed.
System initialize
Use this function to reset DT220 parameters to factory default values. CALIBRATION
FACTORS WILL BE ERASED.
Database initialize
Use this function to erase all ID records.
System & Database initialize
Use this function to reset DT220 parameters to factory default values and erase all ID
records.
Download New Firmware
Use this function to update DT220 firmware. This function should be used by an
authorized IDS dealer or the factory only.
Set Time and Date
When the DT220 is in idle mode (not configuration mode) press the Shift key and then
the function Time Date key use 24 hour time to set the hour, i.e. if it is 3:40 pm enter 15
for the hour. Use the Cont.(↓
↓↓
↓) and Review(↑
↑↑
↑) keys to position the edit block to a time
value. Use the numeric keypad to enter a new value. Press the Enter key to complete the
entry. Position the edit block to Set Time and then press the Enter key to set the clock.
Press Function key Time Date to display time and date for 10 seconds. Press the Shift
key and then Function F4 key to display time in large character format.
NOTE: The clock speed can be adjusted in the Event Monitor section of the
configuration menu.
Kommentare zu diesen Handbüchern